Setting the teachpoints
Teaching part 2: Adjusting teachpoints for accessories
Adjusting the teachpoint at deck location 2 for the Cartridge & Tip Seating Station
66
8
Click the Move to safe height at current X/Y position to remove the probes from
the wash station chimneys, and return the head to the safe Z- height.
9
In the Profiles tab, click the Update this profile to save the new teachpoints.
You adjust the teachpoint for deck location 2 to enhance the precision of the
probes as they move into the cartridges or pipette tips in the Cartridge & Tip
Seating Station.
IMPORTANT
Ensure that no pipette tips or cartridges are seated on the
Bravo 96AM Head. Ensure that the 250- µL tip remains selected as the
Teaching tip type in the Profiles tab.
To adjust the teachpoint for the Cartridge & Tip Seating Station:
1
Ensure that no labware is on the deck other than the wash station at deck
location 1.
2
In the Bravo Diagnostics Jog/Teach tab, click Location 2.
3
Write down the value that appears in the Z-axis box for location 2. You will
use this value in a later step.
Note: Additional accessories will appear in the image if you have
configured them in this profile.
4
In the Approach height above teachpoint (mm) list, select 50, and then click
Approach.
5
Using care to avoid touching the head probes, place the Cartridge & Tip
Seating Station on the platepad at deck location 2. Ensure that the station
is fully seated on the platepad.
AssayMAP Bravo Platform Installation Guide