Setting up the computer and software
Testing the accessories
54
4
Click Run pumps.
5
Verify that the pumps run. While they are running, verify that the pumps
are spinning in the correct directions:
Top pump should be spinning clockwise.
Bottom pump should be spinning counter- clockwise.
6
Click Stop pumps, and then click OK.
AssayMAP Bravo Platform Installation Guide