Testing the Peltier Thermal Station
AssayMAP Bravo Platform Installation Guide
To test the Peltier Thermal Station:
1
Ensure that the MTC Controller is on.
2
In the Accessory Configuration area, click the Heating Shaking Station icon for
deck location 4, and then click Diagnose accessory. The Bravo Accessory
Diagnostics dialog box opens.
3
In the Bravo Accessory Diagnostics dialog box, verify that the correct device
(CPAC Ultraflat) appears in the display at the top.
4
In the Heat area, set the parameters to test the temperature function:
a
In the Target temp box, type the temperature (°C).
For example, set the CPAC at deck location 4 to 45 °C.
b
Click Start.
c
At the Peltier Thermal Station on the deck, verify that the accessory
operates correctly.
You can also monitor the readout in the Current temperature display.
d
Click Turn Off to stop the heating or cooling, and click OK to close the
dialog box.
Setting up the computer and software
Testing the accessories
55