Manually updating teachpoints in each profile

Workflow for setting teachpoints in each profile manually

Workflow for setting teachpoints in each profile manually
About this topic
Before you start
Workflow for manually setting teachpoints
Step
For this task...
1
Initialize the AssayMAP 1 profile.
2
In the AssayMAP 1 profile, set the teachpoint for
deck location 2 and derive the teachpoints for the
other eight deck locations automatically.
3
Fine tune the derived teachpoints.
4
Normalization profile only. If the Bravo Plate Riser
is installed at deck locations 2 and 6, adjust the
teachpoint at these locations to accommodate the
extra height.
5
GlykoPrep- Plus profile only. If the Peltier Thermal
Station is installed at deck location 4, adjust the
teachpoint to accommodate the red PCR plate insert.
6
Orbital Shaking Station only. If the accessory is
installed at deck location 9, fine tune the teachpoint.
7
Adjust the teachpoint for the 96AM Wash Station.
100
Every device profile in the Protein Sample Prep Workbench must be adjusted
to the correct teachpoints of the specific AssayMAP Bravo Platform. You can
use the following workflow to set the teachpoints for all nine deck locations in
every device profile manually.
Alternatively, for a semi- automated workflow, see
teachpoints in AssayMAP 1 profile" on page
Ensure that all device profiles in the Protein Sample Prep Workbench have
been installed, and that they are configured correctly. See
settings and configuration" on page
CAUTION
Using a profile that is associated with teachpoints from a different instrument
can result in a hardware collision and cause equipment damage. Ensure that you verify the
teachpoints associated with each profile for your specific instrument and Bravo 96AM Head.
"Teaching part 1: Setting
58.
"Verifying the profile
39.
See...
"Initializing the profile that you are
verifying" on page 51
"Teaching part 1: Setting
teachpoints in AssayMAP 1 profile"
on page 58
"Fine tuning the teachpoints" on
page 61
"Adjusting the teachpoint for the
Bravo Plate Riser" on page 101
"Adjusting the teachpoint for the
red PCR plate insert" on page 103
"Teaching the Orbital Shaking
Station at deck location 9" on
page 62
"Teaching the 96AM Wash Station
at deck location 1" on page 63