Installing The Standard Platepads - Agilent Technologies AssayMAP Bravo Installation Manual

Platform
Table of Contents

Advertisement

Installing the platform

Installing the standard platepads

Installing the standard platepads

Before you start
Procedure
20
The platepad has a crosshairs for setting pipettor teachpoints in one corner, as
the following figure (1) shows.
1
Install the standard platepads at the Bravo deck locations specified for your
configuration. The standard configuration uses platepads at deck locations 2, 3,
and 5–8.
1
2
00240
Bravo
Plate Orientation
4
5
7
8
Make sure you have the following:
Platepad
Flathead cap screw, M6 x 40 mm
To install a platepad:
1
With the crosshairs toward the back side of the device, align the pins in
the bottom of the platepad with the holes on the deck.
Make sure the pins are inserted correctly and the accessory sits level on
the deck.
2
Secure the platepad to the deck using the screw provided.
3
Ensure that the setscrews in the platepad tabs are recessed and not
protruding into the plate nest. Use a 1.5- mm hex wrench to back out any
protruding setscrews.
3