Agilent Technologies AssayMAP Bravo Installation Manual page 15

Platform
Table of Contents

Advertisement

AssayMAP Bravo Platform Installation Guide
Figure Bottom screw located at outer hole
5
Ensure the risers are installed securely, and the outside edges of the risers
are flush with the Bravo base.
6
Lift the Bravo Platform up onto the risers using proper lifting techniques.
WARNING
Attempting to lift the Bravo device without assistance could cause
personal injury and result in equipment damage. Request assistance and use proper
lifting techniques when lifting the Bravo device.
Figure Bravo Platform on risers with no deck platepads
Ensure that the instrument and riser assembly does not wobble and that it
sits flat on the table. If the Bravo Platform has any wobble or the risers
have uneven seating, replace or realign and tighten the risers.
Installing the platform
Installing the Bravo risers
7