Verification Workflow; About Installing The Bravo Pendant, Light Curtain, And Shields - Agilent Technologies AssayMAP Bravo Installation Manual

Platform
Table of Contents

Advertisement

Verifying the installation and setup

Verification workflow

Verification workflow

About installing the Bravo pendant, Light Curtain, and shields

80
After the AssayMAP Bravo Platform is set up in the VWorks software, perform
the following steps in the order given to verify the installation and setup.
Step
For this task...
1
Ensure that the Light Curtain,
pendant, and shields are
installed.
2
Open the Protein Sample Prep
Workbench and access the
Utility Library.
3
Run the Startup utility.
4
Test the safety equipment.
5
Run the Syringe Wash utility
for 60 cycles.
6
Perform a mock run of the
Affinity Purification protocol.
7
Run the Syringe Test utility.
8
Run the Shutdown utility.
WARNING
Moving-parts hazards are accessible through the front, rear, and side
openings of the Bravo Platform. To prevent potential injury, ensure that the safety
equipment is installed and in good working condition.
Ensure that the following are installed properly. For instructions, see the
Bravo Platform Safety and Installation
Emergency- stop pendant. Ensure that the pendant and its cable are in
good condition and connected to the Pendant port on the junction box.
Light Curtain. Ensure the lightpost transmitter (TX) and receiver (RX)
cables are connected and that the light beams are aligned correctly.
Bravo front, side and rear shields. Ensure that the accessory tubing and
cabling is routed off the Bravo deck through the access windows in the
side and rear shields.
See...
"About installing the Bravo
pendant, Light Curtain, and
shields" on page 80
"Opening the Workbench and
accessing the utilities" on page 82
"Running the Startup utility" on
page 83
"Testing the safety equipment" on
page 84
"Washing the syringes to prepare
for first time use" on page 86
"Performing a mock run" on
page 90
"Running the Syringe Test utility"
on page 97
"Running the Shutdown utility" on
page 98
Guide.