About installing the AssayMAP head

About installing the AssayMAP head
Before you install the head
8
Ensure that you provide 30.48- cm (12- inch) clearance behind the Bravo
Platform.
IMPORTANT
Make sure that you remove the y- axis shipping stop from the
Bravo head mount. For instructions, see the
IMPORTANT
Before you install any platepads or accessories, use a clean lint-
free cloth to wipe the Bravo deck and ensure that it is clean.
The Bravo 96AM Head has four retractable tip- box stripper pins, two on each
side of the probes, as the following figure shows. Before you install the Bravo
96AM Head, ensure that the tip- box stripper pins are retracted, as the
following figure shows.
Figure Bravo 96AM Head (upside down) with stripper pins retracted
If necessary, perform the following procedure to retract the stripper pins.
For instructions on how to install the Bravo 96AM Head, see the
Platform Safety and Installation
CAUTION
If the tip-box stripper pins are extended, the pins will crash into the wash
station during syringe washing. To prevent potential equipment damage, ensure that the pins
are always retracted during use.
Bravo Platform Unpacking
Guide.