Setting the teachpoints
Teaching part 3: Updating the other profiles based on AssayMAP 1
Before you start
Starting the AssayMAP Teachpoint Update protocol
68
Table Special adjustments automatically performed by Teachpoint Update protocol
Profile
GlykoPrep- Plus
AssayMAP
Normalization
Ensure that you have completed the following procedures:
•
"Teaching part 1: Setting teachpoints in AssayMAP 1 profile" on page
•
"Teaching part 2: Adjusting teachpoints for accessories" on page 62
Ensure that you have the required resources:
•
AssayMAP Teachpoint Update protocol (.pro) installed in the following
location:
C:\VWorks Workspace\Workbench\Protocols\AssayMAP Teachpoint Update
v1p0.pro
•
250- µL pipette tip
IMPORTANT
Ensure that the pipette tip is new and in good condition.
Note: The AssayMAP Teachpoint Update protocol is not designed to work
in simulation mode.
To start the AssayMAP Teachpoint Update protocol:
1
In the VWorks window, click File > Open. The Open dialog box appears.
Deck
Teachpoint adjustments for accessories
location
4
Raises the teachpoint z- axis coordinate by
4.4 mm to accommodate the height of the
Red PCR Insert on the Peltier Thermal
Station.
2, 6
Raises the teachpoint z- axis coordinate by
28.5 mm at these deck locations to
accommodate the height of the Bravo Plate
Riser.
AssayMAP Bravo Platform Installation Guide
58.