Teaching part 3: Updating the other profiles based on AssayMAP 1

AssayMAP Bravo Platform Installation Guide
1
00240
Bravo
Plate Orientation
4
7
6
Carefully, use the Jog Axes controls to jog the head down incrementally to
the Cartridge & Tip Seating Station.
7
When the syringe probes are just above the wells, use the Jog Axes controls
for the x- and y- axes to position the probes so that they are perfectly
centered within the openings of the corresponding wells in the Cartridge &
Tip Seating Station.
8
When the probes are perfectly centered within the well openings:
a
Click Teach to save these values temporarily.
b
Click Move to safe height at current X/Y position to return the head to a safe
z- axis height.
9
Remove the Cartridge & Tip Seating Station from the deck.
10
Carefully, use the Down +Z control to jog the head down incrementally until
the value in the Z-axis box matches the value that you noted in step 3.
11
Click Teach to set the new teachpoint.
12
In the Profiles tab, click Update this profile.
The AssayMAP Teachpoint Update protocol will guide you through the process
of setting teachpoints for all other profiles in the Protein Sample Prep
Workbench based on the AssayMAP 1 profile.
Several of the profiles included in the Protein Sample Prep Workbench require
special teachpoint adjustments for accessories that will be placed on specific
Bravo deck locations, which are listed in the following table. The AssayMAP
Teachpoint Update protocol automatically makes these teachpoint adjustments.
Teaching part 3: Updating the other profiles based on AssayMAP 1
2
3
5
6
8
9
00240
Setting the teachpoints