Verifying the installation and setup

Running the Shutdown utility

Running the Shutdown utility

Workflow
98
You should run the Shutdown utility after every application protocol run, if the
AssayMAP Bravo Platform will not be in use for 1 hour or longer, and before
shutting down the device overnight.
The Shutdown utility prepares the AssayMAP Bravo Platform for idle time by:
Washing the syringes the specified number of wash cycles.
Aspirating 200 µL of Syringe Storage Liquid into the syringes. The
Syringe Storage Liquid prevents residual reagents from forming salts
and corroding the syringe seals.
To run the Shutdown utility:
1
Open the Utility Library, locate the System Startup/Shutdown utility, and then
click Utility.
2
Click Instructions.
3
Follow the instructions to run the Shutdown utility.
AssayMAP Bravo Platform Installation Guide