Before you start
Installing the Orbital Shaking Station
AssayMAP Bravo Platform Installation Guide
For a full description of the Orbital Shaking Station, see the manufacturer's
documentation.
Make sure you have the following:
Orbital Shaking Station and the control module
Adapter pad (supplied with the Orbital Shaking Station)
Hex drivers
M6 x 12- mm flathead screw that holds the adapter pad to the deck
(supplied with the Orbital Shaking Station)
Two M3 x 18- mm socket- head cap- screws that attach the Orbital Shaking
Station to the adapter pad (supplied with the Orbital Shaking Station)
Use the following procedure to install the Orbital Shaking Station at deck
location 9. Ensure that no platepad is installed at deck location 9. If necessary,
uninstall the platepad by removing the M6 screw in the center of the platepad.
Figure Deck locations (top view)
1
2
00240
Bravo
Plate Orientation
4
5
7
8
To install the Orbital Shaking Station:
1
Position the adapter pad at the designated deck location, aligning the
dowel pins with the corresponding holes in the deck.
Make sure the adapter pad sits level on the deck.
2
Insert the M6 x 12- mm flathead screw into the center of the adapter pad
and tighten the screw.
3
Place the Orbital Shaking Station atop the adapter pad so that the cable is
towards the back of the Bravo deck.
Make sure the station is completely centered on top of the adapter pad
and sits level.
Installing and connecting the Orbital Shaking Station
3
6
9
00240
Installing the platform
27