Procedure
AssayMAP Bravo Platform Installation Guide
To prime and wash the syringes:
1
Set up the Wash Buffer Reservoir as follows:
a
Fill one of the following labware with the corresponding amount of
deionized water.
Labware
Reservoir, Seahorse 201254- 100, PP, no walls,
pyramid bottom
12- Column, Low- Profile Reservoir
8- Row Low- Profile PolyPro Reservoir
b
Place the filled labware on Bravo deck location 4.
2
Ensure that the deck is clear of other labware, except the wash station at
deck location 1 and the 96AM Cartridge & Tip Seating Station at deck
location 2.
Make sure the 96AM Cartridge & Tip Seating Station empty. (The station
should not contain cartridges or pipette tips.)
3
Open the Syringe Wash utility
Verifying the installation and setup
Washing the syringes to prepare for first time use
Minimum volume (mL)
50 mL
6 mL/column
8 mL/row
.
87