AssayMAP Bravo Platform Installation Guide
5
Carefully, use the Jog Axes controls to position the syringe probes so that
they are perfectly centered over the corresponding chimneys in the wash
station.
The tips of the probes should be approximately 1 mm above the tops of the
chimneys.
6
When the probes and chimneys are perfectly aligned, use the Jog Axes
controls to move the head down slowly along the z- axis to the original
Z-axis teachpoint value that you recorded in step 3.
Note: This step will insert the probes into the wash station chimneys,
which is a normal part of the teaching process.
7
When the position of the z- axis (as indicated by the Z- axis readout in the
bottom right of the screen) matches the value that was written down from
above, click Teach to set the new teachpoint for deck location 1.
Teaching part 2: Adjusting teachpoints for accessories
Setting the teachpoints
65