Installing the platform
Installing the 96AM Wash Station and Pump Module
Installing the 96AM Wash Station
12
Before you start
WARNING
To prevent potential injury, turn off the Bravo Platform before you install or
remove any accessory.
Ensure that you have the following:
96AM Wash Station and the platepad with brackets
Pump Module 2.0
Tubing kit with inline filter
Hex driver for 2- mm screws
Ensure that you install the 96AM Wash Station at deck location 1.
Figure 96AM Wash Station installed at deck location 1
To install the wash station:
1
At deck location 1, remove the standard platepad, if installed, and then
install the bracketed platepad. Use the supplied screw to secure the
platepad to the deck.
Ensure the platepad crosshairs are oriented towards the northwest corner
of the deck, as the following figure shows.
AssayMAP Bravo Platform Installation Guide