Table 4-1
Test equipment required for diagnostic test
Table 4-2 shows the equipment required for performing the diagnostic test.
Table 4-2
Required equipment
Required test equipment
Coaxial cable with N type (m) connectors,
61 cm (24 in)
Calibration kit
Type-N coaxial Short termination
Power splitter
Handler I/O test kit
To Execute the Diagnostic Test
To isolate faulty board assembly in analog section, execute the diagnostic test in
accordance with the following procedure. The test procedure needs to be performed using a
mouse and an external keyboard in addition to the front panel keys.
NOTE
To perform the diagnostic test properly, the following conditions must be met:
1. Environmental temperature: 23C 5C
NOTE
Do not operate front panel keys, keyboard and mouse during the diagnostic test. Changing
the instrument settings while the diagnostic test is in progress will cause incorrect test
results.
Step 1. Connect a mouse and an external keyboard to the E5061A/E5062A's rear panel connector.
Step 2. Press
Step 3. Press
shown in Figure 4-10.
Chapter 4
Diagnostic test group menu
8
Receiver Compression Test
9
Receiver IF Ranging Test
10
Handler I/O Test
key.
Load Project
to select
Load Project
Troubleshooting Using Diagnostic Test
Qty
Recommended model
N6314A, 8120-8898(for 75 )
1
85032B/F, 85036B(for 75 )
1
2
part of 85032B/F, 85036B(for 75
)
0955-0751, 0955-0752(for 75 )
1
1
E5070-65001
function."Open" dialog box will be displayed as
Troubleshooting
101