Test Equipment Required For Diagnostic Test; To Execute The Diagnostic Test - Agilent Technologies E5061B Service Manual

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Table 2-1

Test equipment required for diagnostic test

Table 2-2 shows the equipment required for performing the diagnostic test.
Table 2-2
Required equipment
Required test equipment
24 inch 50Ω cable

To Execute the Diagnostic Test

To isolate faulty board assembly in analog section, execute the diagnostic test in
accordance with the following procedure. The test procedure needs to be performed using a
mouse and an external keyboard in addition to the front panel keys.
NOTE
To perform the diagnostic test properly, the following conditions must be met:
1. Environmental temperature: 23°C ± 5°C
NOTE
Do not operate front panel keys, keyboard and mouse during the diagnostic test. Changing
the instrument settings while the diagnostic test is in progress will cause incorrect test
results.
Step 1. Connect a mouse and an external keyboard to the E5061B's rear panel connector.
Step 2. Press
Step 3. Press
shown in Figure 2-13.
Figure 2-13
Open dialog box
Step 4. Select "User [D]" (preset state) from menu in the "Look in:" box.
Step 5. Double-click "Agilent" folder to open it and to access its menu.
Chapter 2
Diagnostic test group menu
8
Gain-Phase Absolute Measurement Test
(Option 3L5)
key.
to select
Load Project
Load Project
Troubleshooting Using Diagnostic Test
Qty
Recommended model
1
8120-8862, 8120-1839
function."Open" dialog box will be displayed as
Troubleshooting
31