Test Record For Power Meter Accuracy Test - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

General Information
Power Meter Accuracy Test

TEST RECORD FOR POWER METER ACCURACY TEST

Power Meter Tested
Model Number:
Serial Number:
Test Equipment Used
Range Calibrator: Model No.:
Digital Voltmeter: Model No.:
Test Results
Range
Channel A
Calibrator:
DVM Reading
Setting
(Vdc)
1 mW
A=
300 μW
B =
100 μW
D =
Pass/Fail Calculation
300 μW
R=B/A =
(C/10)
S=
10
% ERROR=((R-S)/R)x100 =
Limits: ±0.13%
Pass __ Fail __
100 μW
T=D/A =
(E/10)
U=
10
% ERROR=((T-U)/T)x100 =
Limits: ±0.10%
Pass __ Fail __
26
Test Date:
Tested by:
Power Meter
Reading (dBm)
0.00±0.001
C =
E=
=
=
Serial No.:
Serial No.:
Channel B
DVM Reading
Power Meter
(Vdc)
Reading (dBm)
A=
0.00±0.001
B =
C=
D =
E =
R=B/A =
(C/10)
S=
=
10
% ERROR=((R-S)/R)x100 =
Limits: ±0.13%
Pass __ Fail __
T=D/A =
(E/10)
U=
=
10
% ERROR=((T-U)/T)x100 =
Limits: ±0.10%
Pass __ Fail __
Chapter 1

Advertisement

Table of Contents
loading

Table of Contents