Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
E8362C, E8363C, E836C

Phase Lock IF Gain Adjustment

The phase lock IF gain adjustment is used to adjust the phase-lock loop IF signal amplitude to a constant
level across the analyzer's full frequency range:
1. No additional equipment is required for this adjustment.
2. No connections to the test ports are required for this adjustment.
Procedure
1. On the System menu, point to Service, Adjustments, and then click IF Gain Adjustment.
2. If appropriate, select any special test options installed. Otherwise, leave this selection at None. The
Advanced screen is for factory use only.
3. Click Begin Adj.
4. The adjustment takes approximately one minute to complete.
3-46
Service Guide E8364-90038