Tests and Adjustments
Adjustments
Phase Lock IF Gain Adjustment
The phase lock IF gain adjustment is used to adjust the phase-lock loop IF signal
amplitude to a constant level across the analyzer's full frequency range:
1. No additional equipment is required for this adjustment.
2. No connections to the test ports are required for this adjustment.
Procedure
1. On the System menu, point to Service, Adjustments, and then click IF Gain
Adjustment.
2. If appropriate, select any special test options installed. Otherwise, leave this selection
at None. The Advanced screen is for factory use only.
3. Click Begin Adj.
4. The adjustment takes approximately one minute to complete.
3-48
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Service Guide E8364-90026