Calibration and Adjustment

Aborting a Calibration in Progress

Sometimes it may be necessar y to abor t a calibration after the procedur e
has already been in itiated. You can abor t a calibration at any time by
tur ning off the power. You can abor t a calibration by issuin g a r emote
in ter face device clear message.
CAU T IO N

Adjustment Procedures

You can adjust both the gain and offset for each r an ge on ever y chann el.
Option ally, you can choose to adjust only the offset for each r ange on
each chann el. A full adjustmen t r equir es you per for m both the gain and
offset adjustmen ts for ever y r ange on ever y chann el.

Gain and Offset Adjustment

1
2
3
4
68
If you abort a calibration in progress when the digitizer is attempting to write new
calibration constants to EEPROM, you may lose all calibration constants. T ypically,
upon re applying power, the digitizer will report an error Y ou should not use the
digitizer until a complete re adjustment has been performed. A list of the possible
calibration errors is given on
Read
Test Considerations
Conn ect the DMM to the Calibr ation Output. Set the DMM as follows:
Hi-Z in put
10 NPLC
Aut ozer o On ce
Man ual Tr igger Mode
Make a conn ection to the digitizer using the web inter face (see
page
35).
En ter the Remote Control Page and select the
page
79.
on page 48.
Agilent L4532A/ 34A Service Guide
tab.
Utilities