Adjustment Program

The adjustment program listed in Table 1-1 is required for the adjustments. Using the program,
the computer controls the 4349B and the other pieces of equipment to obtain the adjustment
data, and stores the data in the 4349B's internal EEPROM, A2U1.
The adjustment program runs on the computer listed in Table 1-1. The program should be
installed into your computer according to the procedure described in \Program Installation". To
start the program after it has been installed, refer to \Program Execution".
Figure 2-1 shows the adjustment program ow. The following lists the brief explanation for
each menu item. At the end of each adjustment, the program writes the obtained adjustment
data both to the 4349B internal EEPROM and to the unit-serial-number dependent le on the
disk.
INITIAL SETUP
Channel 1 Adjustment
Channel 2 Adjustment
Channel 3 Adjustment
Channel 4 Adjustment
Rewrite EEPROM
Figure 2-1. Adjustment Program Flow
de nes the standards' calibration values. Update the calibration
values as necessary.
obtains all the channel 1 related data such as input resistance and
scaling. This item should be performed if the performance test fails on
this channel.
performs the same as the \Channel 1 Adjustment" for the channel 2.
performs the same as the \Channel 1 Adjustment" for the channel 3.
performs the same as the \Channel 1 Adjustment" for the channel 4.
writes the adjustment data, which is stored in a unit-serial-number
dependent le, to the EEPROM. Only the adjustment data of
previously adjusted channel(s) will be written.
Adjustments
2-3