15. The Temperature calibration complete message indicates that the
process has been successfully completed.
Figure 113: Calibration: Temperature calibration successfully completed
You may tap Re-Calibrate to repeat the calibration process.
4. 6. 3. Temperature Calibration Lookup
You may look up the current calibration temperature at any time:
1.
Tap the Settings icon from the menu bar, followed by the Calibration
button further down on the Temp Calibration screen that appears
2.
The Temp Calibration screen shown in
screen shows the current Shaker Calibrated Temperature.
NOTICE Alternatively, temperature controlled shakers let you look up the
calibration temperature in the screen where you set the temperature. An
example is shown in the section
4. 7. Firmware Installation
Tapping the Firmware button from the Files and Info screen (see
Info" on page
79) requests you to enter the aministrator passcode, then
opens the Firmware screen shown in
new firmware from an USB memory stick.
You may receive a download link for new firmware from your field service
technician, depending on your service agreement. Be sure to install firmware
from authorized sources only. Software from non-authorized sources may
void the warranty granted by Thermofisher Scientific.
1.
Download the ZIP file from the link provided to you by your field service
technician.
2.
Extract the ZIP file to a blank USB drive.
NOTICE Make sure the USB drive has sufficient space to allow the ZIP file to
be extracted.
Firmware Installation
"Set Temperature" on page 65.