WARNING Always lift the shaker from both sides. Never lift the shaker from the
front or back sides. The shakers are heavy (please refer to
on page 12).
Assign a minimum of two persons to lift and carry a shaker.
CAUTION Never grasp and carry a temperature controlled shaker by the handle of
the transparent hood. The handle may break and the shaker may fall to the ground,
inflicting serious injury to persons.
2. 4. Leveling
The shaker cannot be leveled by itself. The shaker requires a leveled
supporting surface for proper setup.
NOTICE Do not put any shims or other flat objects between the shaker feet and the
supporting surface to level the shaker.
2. 5. Mains Connection
The shaker requires a power source that meets its specifications. Power
supply cables are supplied. Check
WARNING Damage from wrong mains supply or power supply plug. Make sure
that the shaker is plugged only into sockets which have been properly grounded.
Do not operate the shaker with a damaged or an inadequately rated power supply
cable.
NOTICE Electromagnetic radiation may cause interference on the display. This will
not damage, restrict the function of or alter the device. To avoid interference from
electromagnetic radiation, do not bring mobile devices, such as cell phones, into
the direct vicinity of the device. Do not operate the device with other devices of
high electrical power on a common circuit. Do not operate multiple devices on a
common power strip.
To connect the shaker to the power supply follow this procedure:
1.
Turn off the power supply switch located on the right side.
2.
Make sure that the cable specification agrees with the safety standards
of your country.
Leveling
Figure 7 on page 47
53 / 143
"Technical Specifications"
for details.