Thermo Scientific Solaris Series Instruction Manual page 10

Shaker
Table of Contents

Advertisement

Damage to health from shaking explosive or
flammable materials or substances.
Do not shake explosive or flammable materials or
WA R N I N G
substances.
Burnings from hot surfaces.
In a temperature controlled shaker the surfaces like
platform and accessories can get hot under the hood.
C A U TI O N
Do not touch hot surfaces. Wait until the hot surfaces
have cooled down.
Cutting injuries from glass shards.
A disengaged platform and accessories, like clamps,
can lead to broken glass if vessels fall off the shaker.
C A U TI O N
Make sure that the platform and accessories are
installed properly by using the correct tools and
screws. Make sure that the clamps are capable to hold
the load of the vessel and sample at chosen speed.
Clamps with your individually configured load may
have lower speed capability than the stability limitation
of the shaker. Refer to
on page 117
Pay attention to any unusual sound. This can indicate a
disengaged platform or accessories.
Damage to shaker or malfunction due to a
damaged touchscreen.
ƒ
C A U TI O N
ƒ
.
Do not operate the shaker if the touchscreen is
damaged.
Power off the shaker. Disconnect the mains plug.
Have the touchscreen replaced by an authorized
service technician.
9 / 143
"3. 4. Loading and Normal Use"