Thermo Scientific Solaris Series Instruction Manual page 114

Shaker
Table of Contents

Advertisement

Figure 98: Adhesive Mat: Push or Pull Vessel Off Carefully
CAUTION Push or pull gently without much physical force! Vessels may come
loose suddenly and spill their content!
4.
If a vessel refuses to come off, pour a little water around its base.
Allow the mat to air dry before using it again.
Remove the Mat
1.
To release the adhesive mat, lift it on one side and pull it slowly away from
the platform, as shown in
Figure 99: Adhesive Mat: Lift and Pull Mat Away to Remove
Cleaning and Disinfection
If the stickiness diminishes, clean the mat.
CAUTION Do not use any solvents, detergents or cleaning agents other than mild
dishwashing soap without skin care additives. Recommended concentration is max.
2 ml of detergent per 1 liter of water.
NOTICE Direct contact of the mat with concentrated detergent reduces stickiness.
In this case rinsing alone is not enough. If detergent is used directly, rinse the mat,
soak it in water for 24 hours and then dry it for 24 hours to regain full adhesive
power.
NOTICE Disinfect with isopropanol 70% or ethanol 70% only.