Alarms and Alerts
The Alarms and Alerts screen lets you determine how and when alarms and
alerts are being issued.
Figure 40: Set Alarms and Alerts Preferences
Volume: This control lets you change the speaker volume of the tone
that sounds when an alarm or alert is issued. Drag the slider to the left to
decrease or right to increase the volume, then tap Save. This volume setting
will affect both alarms and alerts.
Figure 41: Set Speaker Volume for Alarms and Alerts
High Temperature Alarm / Low Temperature Alarm (temperature
controlled shakers only): These two options let you set a high and a low
temperature alarm threshold relative to the temperature setpoint (see
Temperature" on page
chamber falls below the low or surpasses the high temperature threshold,
the shaker issues an alarm. Choose a temperature threshold value and tap
Save.
Figure 42: Set High and Low Temperature Alarm
NOTICE The alarm thresholds shift accordingly whenever you change the
temperature setpoint.
High Temperature Shutdown / Low Temperature Shutdown
(temperature controlled shakers only): These two options let you set a
high and a low temperature safety shutdown threshold relative to the current
temperature setpoint (see
temperature inside the shaker's chamber falls short of the low or surpasses
the high temperature threshold, the shaker shuts down automatically to
Graphical User Interface
65). Whenever the temperature inside the shaker's
"Set Temperature" on page
71 / 143
"Set
65). Whenever the