Thermo Scientific Solaris Series Instruction Manual page 112

Shaker
Table of Contents

Advertisement

Speed Range and Runtime Limit for Glass Vessel (30% fill)
Limitations for speed and runtime at room temperature for glass vessel (30
% fill). Raised temperature reduces the stickiness. Process of changing
the temperature can release the vessels from the mat. Start shaking with
acclimatized vessels.
Microplate /
DeepWell
96 x 2 ml
125 rpm
no runtime limit
150 rpm
72 h
200 rpm
72 h
250 rpm
72 h
300 rpm
350 rpm
400 rpm
Table 54: Low Adhesion Mat - Speed Range and Runtime Limit at Room Temperature
Microplate /
DeepWell
96 x 2
ml
125
no runtime limit
rpm
150
72 h
rpm
200
72 h
rpm
250
72 h
rpm
300
72 h
rpm
350
48 h
rpm
400
24 h
rpm
Table 55: High Adhesion Mat - Speed Range and Runtime Limit at Room Temperature
Accessories
Beaker
25 ml
125 ml
72 h
72 h
72 h
72 h
Erlenmeyer
25
125
250
ml
ml
ml
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
48 h
48 h
48 h
24 h
24 h
24 h
111 / 143
250 ml
72 h
72 h
500
1000
2000
ml
ml
ml
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
72 h
48 h
48 h
48 h
24 h
24 h
500 ml
72 h
72 h
3000
5000
ml
ml
72 h
72 h
72 h
72 h
48 h
24 h
24 h

Advertisement