Thermo Scientific Solaris Series Instruction Manual page 139

Shaker
Table of Contents

Advertisement

No.
Description
5
Drive acceleration too slow.
Desired set speed cannot be
reached.
7
Speed measurement on Fan 1
reports unexpected standstill.
8
Speed measurement on Fan 2
reports unexpected standstill.
9
Speed measurement on Fan 3
reports unexpected standstill.
10
Speed measurement on Fan 4
reports unexpected standstill.
12
Drive start up error - no speed
signal received.
24
Speed measurement error
during a run
26
Speed measurement detected
unexpected standstill during
run.
82
Motor current measurement
detects overload
83
Motor current measurement
out of boundaries
Table 59: List of Errors, Alerts and Alarms
Shipping and Disposal
Solutions
The desired set speed could not be
reached within time. Please check your
settings and/or load (accessories and
samples) on the platform.
Please power-cycle the unit. If the error
message persists, call a service technician.
Please power-cycle the unit. If the error
message persists, call a service technician.
Please power-cycle the unit. If the error
message persists, call a service technician.
Please power-cycle the unit. If the error
message persists, call a service technician.
Shaker platform is blocked. Please check
if enough space is available around
platform and/or reduce load (accessories
and samples) on platform. Afterwards
press START again on the shaker. If the
error message persists, contact a service
technician.
Abnormal speed change detected. Please
check load (accessories and samples)
and/or clamps on the shaker platform.
Afterwards press START again on the
shaker. If the error message persists,
contact a service technician.
Make sure that there is enough space is
available around the platform and/or reduce
load (accessories and samples) on the
platform.
Make sure that the fuse on the back of the
shaker has not tripped
47).
Afterwards press START again on the
shaker. If the error message persists,
contact a service technician.
Motor overcurrent detected. Do not load
or unload platform while running. Reduce
speed or adjust load (accessories and
samples) on platform.
Motor overcurrent detected. Do not load
or unload platform while running. Reduce
speed or adjust load (accessories and
samples) on platform.
138 / 143
Icon
("Fuse" on page

Advertisement

Table of Contents
loading

Table of Contents