Thermo Scientific Solaris Series Instruction Manual page 130

Shaker
Table of Contents

Advertisement

2.
Wait for the Temp Calibration screen shown in
Figure 107: Calibration: Start screen
Tap the Start Calibration button to get started.
3.
Figure 108: Calibration: Prompt for filling Erlenmeyer flask
4.
Follow the instructions on the screen: Fill the 250  ml Erlenmeyer flask
with 100 ml of water.
5.
Install the flask in a clamp and secure the clamp at the approximate
center of the shaking platform.
6.
Suspend the probe of a temperature measuring device of known
accuracy into the flask. The probe should be submerged in the liquid but
not in contact with the bottom or sides of the flask.
7.
Seal the flask to avoid spilling of liquid.
8.
Close the hood over the shaking platform.
Temperature Calibration
129 / 143
Figure 107
to appear.