NOTICE If a pop up screen appears informing you that
temperature may be out of range due to the current ambient temperature, the
ambient temperature is outside of the shaker's operating temperature range,
as specified in the chapter
that you have been notified and proceed, or select a different temperature.
NOTICE The Temp Setpoint screen presents a readout of the calibrated
temperature and offset adjusted during calibration, as shown in the lower
Figure
left section of
Calibration" on page
To return to the main screen, acknowledge the Your Settings have
4.
been saved confirmation dialog that appears.
NOTICE Temperature controlled shaker have a specific behavior when achieving the
set temperature called temperature overshoot. Temperature overshoot means that
the set chamber temperature will first be overdriven to a slightly higher (or lower)
temperature and then approaching the set temperature. While approaching the set
temperature the touchscreen will not show this overshooting behavior. Instead it
will show the rising (or falling) of the chamber temperature until reaching the set
temperature.
Start and Stop the Shaker
1.
To start the shaker, tap the Start button.
The Start button turns into the Stop button.
Figure 34: Start and Stop Button
2.
To stop the shaker, tap the Stop button.
Graphical User Interface
"Technical Data" on page
33. The calibration process is described in
128.