NOTICE Before placing vessels on the platform assembly, make a final check to be
sure that the platform assembly does not wobble.
Figure 80: Dual Stack Platform on a Solaris 4000 Shaker
1.
Carefully position the assembled dual stack platform horizontally over the
shaker and its platform mounting plate.
Platforms for Solaris 2000 have 3 mounting holes, Platforms for Solaris
4000 have 4 mounting holes.
2.
Tighten the screws gently to secure the platform to the shaker. When the
locking tool starts bending, stop tightening.
3. 3. 2. Flask Clamp and Vessel Installation
CAUTION Biological harm due to broken or leaking vessels. Improperly installed
accessories can lead to spilled samples. Make sure that accessories are installed
properly by using the correct tools and screws. The installation requires a
screwdriver (PH2) with a shaft length of 150 mm / 6 in. (Article Number 75004131).
Make sure that accessories fit reasonably on the platform. Always use a vessel
with an accessory of the proper size. Vessels must be intact and installed properly.
CAUTION Cutting injuries from sharp edges. Be cautious while handling platforms
and other accessories.
CAUTION Safe fixation of a clamp is only ensured in combination with the provided
screws. Install the provided screws in all designated mounting bores of that clamp.
NOTICE If the screws are too long, the stability and functionality of the
interchangeable platform may be impaired. Make sure to use only the screws
included in the delivery for fastening.
Accessories
94 / 143