Thermo Scientific Solaris Series Instruction Manual page 134

Shaker
Table of Contents

Advertisement

3.
If still running, stop the shaker by pressing the Stop button on the display
window.
4.
Acknowledge any alarms that may be present, as explained in the
section
"Status" on page
NOTICE If you try to upgrade while the shaker is running, you will receive a
message saying that An upgrade of the shaker is not allowed while the
shaker is actively shaking� That message has a Stop Shaker button to let
you exit and carry out the upgrade.
5.
Tap the Files and Info icon from the navigation bar.
Scroll the Files and Info screen all the way down to reveal the
6.
Firmware button.
7.
Tap the Firmware button to open the Firmware screen showing the
currently installed firmware versions.
8.
Insert the USB drive into any of USB port of the shaker. The location
of the USB ports can be taken from the section
page 44.
9.
Enter the administrator passcode.
Figure 114: Firmware Screen
10. Tap the Upgrade Firmware button.
NOTICE If the shaker cannot find a USB drive, you will be prompted to Please
insert a USB stick with the firmware you like to install to begin the
upgrade� Insert the USB drive now or check whether it is properly plugged into
the USB port and corrrect the problem.
Firmware Installation
67.
133 / 143
"Product Overview" on