Thermo Scientific Solaris Series Instruction Manual page 131

Shaker
Table of Contents

Advertisement

9.
Tap the Next button to continue.
Figure 109: Calibration: Prompt for starting calibration
NOTICE The shaker starts shaking immediately as soon as you tap the Start
button.
10. Follow the instructions on the screen: Tap the Start button to use the
preset temperature, or tap the Temperature Setpoint field and enter
any other desired temperature before tapping the Start button.
The screen shows how the shaker chamber is heated (or cooled) to the
selected calibration temperature.
Figure 110: Calibration: Chamber stabilizing to calibration temperature
Temperature Calibration
130 / 143