Normal Use

Information on speed-load limitation is given only as a guide to product use.
Smooth operation with low or no vibration depends on multiple factors as fill
level, type and condition of clamps and mats, type of vessel, arrangement
of vessels on platform, the surface the shaker is placed on and on the set
speed.
Whenever setting up a new or unknown combination or whenever any
of these parameters change: Increase speed above 175 rpm slowly and
check for unwanted load vibration or movement of the shaker. It is your
responsibility to operate the shaker safely.
Solaris shakers are specially designed for low heat output and low vibration,
which makes them highly suitable for use inside a range of lab equipment,
including environmental chambers, incubators and lab refrigerators. Due to
various specific environmental conditions including the stability of shelves
and supports the user has the responsibility for safe operation when used
inside any lab equipment. Increase speed slowly and check for unwanted
instrument movement.
Safe Speed
The load-independent safe speed for Solaris shakers can be taken from
Table
56. The load includes the weight of the platform, accessories and
samples installed on the shaker. Refer to
details on maximum load.
Model
Solaris 2000
Solaris 4000
Solaris 2000 I
Solaris 2000 R
Solaris 4000 I
Solaris 4000 R
Table 56: Maximum safe speed
Loading and Normal Use
Max� Safe Speed
175 rpm
175 rpm
250 rpm
250 rpm
250 rpm
250 rpm