11. Wait at least two and a half hours (150 minutes) for the chamber
temperature measurement to reach a stable state.
A stable state is indicated by the message Chamber temperature has
stabilized, as shown in
Figure 111: Calibration: Chamber successfully stabilized to calibration
temperature
12. Follow the instructions on the screen:
»
As soon as the screen shown in
the temperature reading of your thermocouple regularly. If you
deem the temperature stable you can either tap the Edit button
to correct or tap the Finish button to complete the calibration
process.
Figure 112: Calibration: Correct temperature displayed by shaker
13. Using the wheel picker that appears, correct the temperature displayed
by the shaker to the reading of the accurate measuring device.
14. Tap the Save button to have your changes registered.
Temperature Calibration
Figure 111
below.
Figure 111
131 / 143
is shown, compare