2. 2. Location
CAUTION Protection can be impaired due to reduced stability of plastic exposed to
ultraviolet rays. Do not subject the shaker and plastic accessories to direct sunlight
or other sources of ultraviolet rays.
Put the shaker on a level table or bench capable of supporting the weight
of the shaker with any accessories and samples while in operation.
Place the shaker near an electrical outlet that matches the nameplate
requirements. Allow clearance around the unit for free air convection,
accessory attachments and user convenience.
Mind the following requirements for setting up the unit:
To ensure sufficient clearance for moving parts and avoid squeezing
ƒ
injuries or damage to adjacent equipment, the open air shaker requires
a clear area of 8 cm (3 in) on all four sides of the platform. The
platform of the open air shaker may overlap the casing.
Shaker cause vibrations. Do not store sensitive devices or dangerous
ƒ
objects or substances in the safety zone.
8 cm (3 in)
8 cm
(3 in)
8 cm (3 in)
Figure 8:
Set Up with Clearance of 8 cm (3 in) for Open Air Shaker
To ensure proper ventilation and optimum performance, the following
spacing requirements must be met for temperature controlled shakers:
Solaris 2000 I and 4000 I incubated shakers require a clear area of 8
ƒ
cm (3 in) on all four sides of the casing, as shown on the left side of
Figure
9.
Solaris 2000 R and 4000 R refrigerated shakers may be operated with
ƒ
a clear area of 8 cm (3 in) on all four sides of the casing if there is free
air space above their hood, as shown on the left side of
If the air space above the Solaris 2000 R and 4000 R refrigerated
ƒ
shaker is obstructed, the clearance around the casing must be
Location
8 cm
8 cm
(3 in)
(3 in)
50 / 143
8 cm (3 in)
8 cm
(3 in)
8 cm (3 in)
Figure
9.