3. 3. 1. Platform Installation
CAUTION A full list of compatible platforms for each shaker appears in the section
"1. 2. 1. Platforms" on page 19.
be operated with the factory pre-installed platforms.
CAUTION Cutting injuries from sharp edges. When uninstalling a platform, grasp
under the platform. Do not lift a platform by clamps mounted on it. Be cautious
while handling platforms and other accessories.
CAUTION Hand or Fingers can be squeezed by a moving 11x14 platform in the
Solaris 4000 I / 4000 R shaker. Do not use the 11x14 platform in the Solaris 4000
I / 4000 R shaker.
CAUTION Use only the screws and the locking tool supplied with the platform.
Using other screws or wrong locking tools will lead to improper installation and
therefore possibly can damage the shaker and accessories.
Always install a platform with all screws.
Always use the proper locking tool:
3/16" locking tool (GT530066) for all platforms of Solaris 2000 and
ƒ
Solaris 2000 I/R and 4000 I/R shakers
7/32" locking tool (GT530080) for all platforms of Solaris 4000 shakers
ƒ
Accessories
Temperature controlled shaker models may only
91 / 143