4� 5� Autoclaving - Thermo Scientific Solaris Series Instruction Manual

Shaker
Table of Contents

Advertisement

NOTICE On temperature controlled shakers, make sure you leave the hood
open so cleaning agents and water can evaporate.
You are responsible that the level of decontamination is achieved according
to your requirements.
4. 5. Autoclaving
NOTICE Never exceed the permitted temperature and duration when autoclaving.
No chemical additives are permitted in the steam.
Before autoclaving:
Disassemble all accessories.
ƒ
Thoroughly rinse off any trace of chemicals or detergents. Include a
ƒ
final rinse with distilled water.
Follow this table to check autoclavability:
Item
Shaker
Platform
Clamp (without spring tubes)
Screw Driver
Microplate / DeepWell Plate Clamp
Adhesive Mat
Beaker Rack
Separatory Funnel Clamp
Test Tube Rack – Metal Parts
Test Tube Rack – Plastic Parts
Test Tube Rack – Foam Inserts
Adjustable Vessel Clamp – Metal Parts
Adjustable Vessel Clamp – Anti-slide Mat
Utility Tray – Metal Tray
Utility Tray – Rubber Mat
Adjustable Angle Test Tube Rack Holder –
Metal Parts
Adjustable Angle Test Tube Rack Holder –
Plastic Parts
Adjustable Angle Test Tube Rack Holder
– Foam Inserts
Table 58: Autoclavability of materials
Make sure that the necessary sterility is achieved according to your
requirements.
Appearance and color may change slightly after autoclaving.
Autoclaving
Autoclavable
No
Yes
Yes
No
Yes
No
Yes
Yes
Yes
No
No
Yes
No
Yes
No
Yes
No
No
127 / 143
Specifications
-
121 °C, 20 min
121 °C, 20 min
-
121 °C, 20 min
-
121 °C, 20 min
121 °C, 20 min
121 °C, 20 min
-
-
121 °C, 20 min
-
121 °C, 20 min
-
121 °C, 20 min
-
-