Thermo Scientific Solaris Series Instruction Manual page 113

Shaker
Table of Contents

Advertisement

Preparation
CAUTION Always use protective goggles and gloves while handling glassware.
Before placing vessels on the adhesive mat:
1.
Check vessels for damage, never use damaged vessels.
2.
Wipe down the vessel with a clean and dry wipe to remove all moisture.
3.
Remove all grease and other contamination.
4.
Check that the vessel bottom and adhesive mat are dry and clean.
CAUTION Even a barely visible film or single drop of water on the bottom of
the vessel may cause the vessel to loose adhesion and come off the mat after
a certain runtime.
Figure 97: Adhesive Mat: Wipe Vessel Dry
5.
Reduce the fill levels or seal the vessels to avoid any spillage of fluid.
CAUTION Even if a vessel is sticking perfectly, a single drop of liquid running
down on the outside wall may cause the vessel to come off the mat after a
certain runtime.
Install Vessels
1.
Place vessels on mat and press down.
2.
Push or pull gently to ensure each vessel sticks reliably to the mat.
NOTICE Bear in mind that condensed water may form at low temperatures or
during a prolonged run, possibly causing accidental release of vessels. Sample
and vessel temperature should not deviate from the specified temperature
range of the shaker. For details on the possible temperature range of your
shaker refer to
"1. 1. Technical Data" on page
Remove Vessels
1.
Press the Stop button on the display window or wait for the run to end.
2.
Wait for the platform to come to a complete standstill.
3.
Pull neck of vessel gently and evenly to the side and wait a few seconds.
Vessels with large bottom surfaces may take a few seconds to release.