2.2.4 Pre-installation Requirements

The Thermo Scientific Matrix PlateMate 2x3, with its compact size and small
footprint requires little bench space. However, additional bench space is needed
for reagents, buffers, waste container, and computer system.
Figure 2-4: Selecting the Proper Location
Consider the following requirements when selecting a place for your system:
Flat surface capable of holding 110 lbs (50 kg) plus weight of computer
Ventilation space: 23 in. (584 mm) height + 29.5in (749 mm) depth + 25 in
(635 mm) width (space around instrument)
Bench space on left side for computer and monitor (6.5 ft. serial cable is
provided)
Bench space on right side for reagent supplies (approx. 1 ft.; 3 ft. tubing
provided)
Grounded wall outlet supporting 110/220 VAC (standard outlet for instrument,
plus outlets for PC system)
PC system for ControlMate software
User Manual
2.2 Instrumentation
29.5"
(749 mm)
(635mm)
23"
(584 mm)
25"
11