7.16 Replacing the Doors

When
Parts required
Follow these steps
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Damage of door
TIP The maintenance procedures do not require that you remove the
doors. If this should ever be required for a specific reason or procedure,
follow the related steps in this section.
Replacement door
NOTICE
To avoid damage to the door hinges, be careful when performing the
following sequence of steps and do not apply force.
1. To remove a door, push the door upward while opening.
Open the door to a position in which the two hinges on the housing
are aligned in the grooves on the door. You can remove the door
only when the hinges are in the grooves.
Figure 84: Unhinging a door
No.
Description
1
Hinge on the housing
2
Reception groove on the door
2. Slightly tilt the door to the outside, away from the housing, and
remove the door.