Measurement Procedure

When making Tuned RF Level measurements, optimal measurement accuracy and
NOTE
speed is obtained by setting the frequency on the PSA directly using the
FREQUENCY Channel
Press the Restart key to optimize the measuring receiver configuration if you make
CAUTION
any changes to the UUT or to the measuring receiver settings after you make your
first measurement.
Using a Power Meter with a N5532A/B Sensor Module (Absolute TRFL
Measurement)
Step 1. Connect, configure and verify the measurement system. See
Connections" on page 44
Step 2. Calibrate the Power Meter see
Step 3. Connect the sensor module to the signal source and assure proper connector care
and connector torque settings are followed. SMA/3.5 mm connectors should
receive 8 inch-pounds of torque and Type N connectors 12 in-pounds.
Step 4. On the PSA, perform the Auto Align routine by pressing
All Now
Step 5. Set the signal source amplitude and frequency. For this example use 0 dBm and 50
MHz.
Step 6. Configure the PSA settings as follows:
Table 3-2
PSA Settings
Parameter
Note: Do not use the Frequency Counter
function. You must press the Frequency
key and set the frequency manually.
Chapter 3
key (do not use the frequency counter).
and
"System Configuration" on page
"Zero and Calibrate the Power Meter" on page
.
Mode
Measure
Frequency
Accuracy
(Under Meas Setup)
IF BW
(Under Meas Setup)
Range 3 Switch Delay
(Under Meas Setup, More)
Early instruments only.
Making Measurements
Tuned RF Level Measurement
"System Hardware
47.
,
System
Alignments
Setting
Measuring Receiver
Tuned RF Level
Same as signal source frequency
High
10 Hz
(default)
ON
53.
,
Align
75