4. Initialize the test equipment parameters as follows:
Table 8-7
Signal Generator Settings
Parameter
Frequency
Level
RF
Table 8-8
PSA Settings
Parameter
Note: Do not use the Frequency Counter function. You
must press the
Frequency
5. Both the signal source and the PSA must warm up for at least 2 hours.
6. Look at the calibration report for the step attenuator. The metrology lab that
characterized the attenuator should have listed on the report a note about which
11713A Attenuator Driver switch was activated to obtain the 40 dB data. This
is important since the 8496G has two 40 dB steps, and you must assure you
know which switch to press to enable the calibrated 40 dB step. The other 40
dB attenuator step is used only to achieve the 80 to 110 dB steps.
7. Using the step attenuator calibration data, fill in
Record,"
Atten Calibration Value and Atten Uncert with the reference
attenuation value (the value the cal lab measured) and the measurement
uncertainty for each attenuator step.
8. Select the 0 dB setting on the step attenuator.
9. Press
,
Meas Setup
Set Ref
key again.
Chapter 8
N5531S Measuring Receiver Performance Verification
Performance Verification Tests
Setting
50 MHz
+10 dBm
On
Mode
Measure
Frequency
key and set the frequency
manually.
Accuracy
(under Meas Setup)
Range 3 Switch Delay
(under Meas Setup, More)
(Early instruments only)
10 MHz Out
(under System, Reference)
Freq Ref
(under System, Reference)
Table 8-9, "Measurement Test
key. Assure reading is 0.000 dB. If not, press
Setting
Measuring Receiver
Tuned RF Level
50 MHz
High
ON
ON
Int
Set Ref
375