Example:
4.4.10.1.4 Ratio Mode
The ratio and previous ratio functions permit any measurement result to be scaled to a reference. The reference
may be either a measurement result or a keyboard entry. The Log/Linear key enables measurement results to be
displayed in linear or logarithmic units.
Mode:
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Dependencies/Couplings:
Example:
4.4.10.2 Meas Setup
Displays the measurement setup menu for the Audio Frequency Measurement when the
has been selected in the
4.4.10.2.1 Avg Mode
Selects the type of termination control used for averaging. This determines the averaging action after the
specified number of average count is reached.
Repeat - After reaching the average count, the averaging is reset and a new average is started
•
Exponential - Each successive data acquisition after the average count is reached, is exponentially weighted
•
and combined with the existing average.
Mode:
Key Path:
Remote Command:
Chapter 4
:DISP:AUDF:RREF 400 Hz
:DISP:AUDF:RREF?
:DISP:AUDF:RREF:AUTO OFF
:DISP:AUDF:RREF:AUTO?
MRECEIVE
Amplitude/Y Scale
:DISPlay:AUDFreq:RMODe LOG |LINear
:DISPlay:AUDFreq:RMODe?
LOG
Saved in instrument state.
Log | Linear
No
:DISP:AUDF:RMOD LOG
:DISP:AUDF:RMOD?
menu.
Measure
MRECEIVE
Meas Setup
[:SENSe]:AUDFreq:AVERage:TYPE REP|EXP
[:SENSe]:AUDFreq:AVERage:TYPE?
Front-Panel Key and SCPI Command Reference
Measurement keys
Audio Frequency
key
255