4.3.8 Modulation Rate
Allows the user to switch to the Modulation Rate measurement.
Mode:
Key Path:
Remote Command:
Notes:
Example:
4.3.8.1 SCPI Remote Commands
:CONFigure:MODRate
:INITiate:MODRate
:FETCh:MODRate[n]?
:MEASure:MODRate[n]?
:READ:MODRate[n]?
Index: n
Results Returned
<Mnemonic>
0
Returns unprocessed I/Q trace data, as a series of comma-separated trace points, in volts. The I
values are listed first in each pair, using the 0 through even-indexed values. The Q values are the
odd-indexed values.
not specified or
Returns comma-separated scalar results, in the following order:
n = 1
Return the measured modulation rate (Hz).
Modulation Rate Ratio (–999 in Normal mode, Ratio: dB or % depending on Ratio Mode setting)
Ratio Reference Frequency (Hz)
Return demodulated result(% for AM, Hz for FM, and rad for PM)
Return modulation distortion(%)
Return modulation sinad(dB)
Chapter 4
MRECEIVE
Measure, More
:CONFigure:MODRate
This key invokes Modulation Rate Measurement.
:CONF:MODR
Front-Panel Key and SCPI Command Reference