Agilent Technologies PSA Series Manual page 205

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Key Path:
Remote Command:
Default:
State Saved:
Range:
Example:
4.4.4.5.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
Key Path:
Remote Command:
Preset/Default:
State Saved:
Min:
Max:
Resolution
Example:
Chapter 4
Meas Setup
[:SENSe]:AMDepth:TRIGger:SOURce
IMMediate|IF|EXTernal[1]|EXTernal[2]|RFBurst
[:SENSe]:AMDepth:TRIGger:SOURce?
IMMediate
Saved in instrument state.
Free Run (Immediate)|Video (IF Envlp)|RF Burst|Ext Front|Ext Rear
:AMD:TRIG:SOUR IF
:AMD:TRIG:SOUR?
Meas Setup
[:SENSe]:AMDepth:CAPTure <time>
[:SENSe]:AMDepth:CAPTure?
[:SENSe]:AMDepth:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:AMDepth:CAPTure:AUTO?
250 ms
Saved in instrument state.
1 μs
10 s
1 ms
:AMD:CAPT 0.25
:AMD:CAPT?
:AMD:CAPT:AUTO OFF
:AMD:CAPT:AUTO?
Front-Panel Key and SCPI Command Reference
Measurement keys
205

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5531sE4440aE4443aE4445aE4446aE4447a ... Show all

Table of Contents