Agilent Technologies PSA Series Manual page 129

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4.2.1.7 Auto Carrier Frequency
Turns the auto carrier frequency estimation in analog demodulation measurements on or off. This key is only
available if PSA firmware revision A.11.08 or above, or Option 23A is installed.
Mode:
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Restrictions and Notes:
Example:
4.2.1.8 Filter Type
This parameter is used to determine the way that the high pass and low pass filters are implemented. the option
for this parameter are the following:
FIR (0dB)
These filters have 0 dB of loss at the stop frequency. The measurement speed with this type of
filter is slow.
IIR (–3dB)
These filters have 3 dB of loss at the stop frequency. The measurement speed with this type of
filter is faster than the FIR filter.
Mode:
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Example:
Chapter 4
MRECEIVE
,
Det/Demod
More
[:SENSe]:MRECeive:CARFreq:AUTO[STATe] ON|OFF|0|1
[:SENSe]:MRECeive:CARFreq:AUTO[STATe]?
ON
Saved in instrument state.
ON | OFF
This key is only applied to analog measurements. It will be unavailable
(grayed out) in the audio measurements.
MREC:CARF:AUTO ON
MREC:CARF:AUTO?
MRECEIVE
,
Det/Demod
More
[:SENSe]:MRECeive:FILTer:TYPE FIR|IIR
[:SENSe]:MRECeive:FILTer:TYPE?
FIR
Saved in instrument state.
FIR (0dB)
IIR (–3dB)
|
MREC:FILT:TYPE IIR
MREC:FILT:TYPE?
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
129