Agilent Technologies PSA Series Manual page 259

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Range:
Example:
4.4.11.2 Meas Setup
Displays the measurement setup menu for the Audio AC Level Measurement when the Audio AC Level
key has been selected in the
Measurement
4.4.11.2.1 Average Mode
Selects the type of termination control used for averaging. This determines the averaging action after the
specified number of frames (average count) is reached.
Exponential - Each successive data acquisition after the average count is reached, is exponentially weighted and
combined with the existing average.
Repeat - After reaching the average count, the averaging is reset and a new average is started.
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Example:
4.4.11.2.2 Average Number
Sets the number that will be averaged. After the specified number (average counts) have been averaged, the
averaging mode (termination control) setting determines the averaging action.
Key Path:
Remote Command:
Preset:
State Saved:
Chapter 4
Log | Linear
:DISP:AUDL:RMOD LOG
:DISP:AUDL:RMOD?
menu.
Measure
Meas Setup
[:SENSe]:AUDLevel:AVERage:TYPE REP|EXP
[:SENSe]:AUDLevel:AVERage:TYPE?
REP
Saved in instrument state.
Repeat | Exponential
:AUDL:AVER:TYPE EXP
:AUDL:AVER:TYPE?
Meas Setup
[:SENSe]:AUDLevel:AVERage:COUNt <integer>
[:SENSe]:AUDLevel:AVERage:COUNt?
[:SENSe]:AUDLevel:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AUDLevel:AVERage[:STATe]?
25, ON
Saved in instrument state.
Front-Panel Key and SCPI Command Reference
Measurement keys
259