Agilent Technologies PSA Series Manual page 251

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

The minimum bandwidth used is determined by the Low Pass Filter setting and is a value that is >
2xLowPassFilter. This mode usually works for most signals with a low modulation index. This mode will
increase measurement throughput because the analyzer can bypass the bandwidth measurement routines that
are performed at initialization for the Auto mode.
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Notes:
Example:
4.4.9.2.3 Average Number
Allows you to specify the number of measurement averages used when calculating the measurement result.
Key Path:
Remote Command:
Preset:
State Saved:
Min:
Max:
Example:
Chapter 4
Meas Setup
[:SENSe]:MODSinad:BANDwidth|BWIDth:TYPE MIN|AUTO|MAN
[:SENSe]:MODSinad:BANDwidth|BWIDth:TYPE?
AUTO
Saved in instrument state.
Auto|Man|Min
The IF BW type "Min" is meaningful only when the Modulation Mode is
"AM". For FM, PM signal, the result will not be correct with this IF BW
Type.
MODSinad:BAND:TYPE AUTO
MODSinad:BAND:TYPE?
Meas Setup
[:SENSe]:MODSinad:AVERage:COUNt <integer>
[:SENSe]:MODSinad:AVERage:COUNt?
[:SENSe]:MODSinad:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MODSinad:AVERage[:STATe]?
25, ON
Saved in instrument state.
1
8192
:MODSinad:AVERage:COUNt 25
:MODS:AVER:STAT OFF
:MODS:AVER:STAT?
Front-Panel Key and SCPI Command Reference
Measurement keys
251